Features

2U cabinet including one TR/Switch for two coils (rat & mouse), one selective preamplifiers and HF filters, 2 fast PIN diode driver for automated transmit/receive switching for the 2 coils (rat/mouse & surface coil), the hybrid coupler for circular polarization and the automated identification and detection of the RF coils.

Domain  MRI NMR
Model number CIB MRI PREAMP NMR
TR switch power 100W pulsed for rat or mouse 150 W
Switching time 40 µs
Coil identification Two slot resistor coding 1: rat/mouse, 2: surface coil /
Auto coil detection yes /
Way name / 1H, X, 2H
Preamp. Noise factor 0.5 dB
Preamp. Gain 26 dB
Hybrid coupler 0.125-0.250 GHz*, 2 way splitter/combiner. 100 W /
Integrated power supply -30Vdc, 5Vdc, 12Vdc, 15Vdc 5Vdc, 12Vdc, 15Vdc
Main power 110-240V 50Hz, single phase

*Specifications depending on the frequency of the magnet